Copper Mountain TR Series 2-Port VNA up to 4.8GHz

min Quantity
1 piece
Unit price
base price
Variants
technical Question?
recommend by email
close
Please fill in all fields.
Question Type
Your E-mail Adress
My Question:
Please validate that you are not a robot.
Description
Accessories
Files/Downloads

Benefits of the Copper Mountain TR1300/1 2-Port USB PC VNAs

  • Dynamic range up to 125  dB.
  • Wide output power range from -50...+5 dBm.
  • Models with 50 or 75 Ω.

Copper Mountain TR Series 2-Port VNA up to 4.8 GHz

The Copper Mountain compact series TR5000/TR7000 VNAs (vector network analyzers) delivers lab grade performance in a compact package, with all the features engineers have come to expect included standard in the software.

These portable VNAs can be battery powered and used in the field, in the laboratory, and in production testing. The devices can be integrated into a production test system via manufacturing test plugin.

The Copper Mountain compact series TR5000/TR7000 VNA includes an RF measurement module and the TR software application which runs on Windows or Linux operating system on a PC, laptop, tablet, or x86 board computer connecting to the measurement hardware via USB interface. The TR software can be installed on multiple computers, making it easy to share the use of the analyzer measurement module.

  • Frequency range up to 1.3 GHz, 3.0 GHz, or 4.8 GHz.
  • Measured parameters S11, S21.
  • Sweep types: Linear frequency, log frequency, segment, power sweep.
  • Impedance 50 Ω or 75 Ω.

Applications

The Copper Mountain Technologies TR compact series VNAs are designet for applications that include RF component design and testing, on-wafer probing, field testing, insertion loss measurement, distance to fault measurements, antenna matching, quality control, material measurements, and many others.

Model Overview

ModelTR1300/1TR7530 - obsoleteTR5048 - obsolete
Ports222
Frequency range300 kHz...1.3 GHz20 kHz...3.0 GHz20 kHz...4.8 GHz
Measured parametersS11, S21S11, S21S11, S21
Sweep typesLinear frequency, log frequency, segment, power sweepLinear frequency, log frequency, segment, power sweepLinear frequency, log frequency, segment, power sweep
Dynamic range (10 Hz IFBW)130 dB, typ. 135 dB75 dB/typ. 100 dB (20...300 kHz)
120 dB/typ. 123 dB (300 kHz...3.0 GHz)
75 dB/tpy. 100 dB (20 kHz...300 kHz)
120 dB/typ. 123 dB (300 kHz...4.8 GHz)
Measurement speed/measurement time per point150 µs250 µs200 µs
Output power adjustment rate-55...+3 dBm-50...+5 dBm-50...+5 dBm
Measurement pointsup to 16,001up to 200,001up to 200,001
Impedance50 Ω75 Ω50 Ω
InterfaceUSB 2.0 (USB type B)
Power supplyExternal power adaptor
Recommended accessoriesAutom. cal. module ACM2506
Cal. kit N1.2, S911T
Cables and adaptors 2x C50SMNM.X, 2x C50NMNM.X
Torque wrench: TW-N
Autom. cal. module ACM4000T
Cal. kit F7511
Cables and adaptors 2x C75NMNM.X, 2x C75NMFM.X
Torque wrench: TW-N
Autom. cal. module ACM2506
Cal. kit N611, N612, S911T
Cables and adaptors 2x C50NMNM.X, 2x C50NMNM.X+ NF50, 2x C50SMNM.X
Torque wrench: TW-S, TW-N

VNAs and ACMs are delivered with factory calibration certificates containing no data. The add-on option for ISO17025/Z540-1 accredited, traceable calibration certificate and uncertainties is available and needs to be specified at time of order.