HIOKI Functional Units for the MR Series Memory Recorders
HIOKI Functional Units for the MR Series Memory Recorders
Most of the HIOKO memory recorders have a high flexibility due to a modular design. Each recorder consits of a main unit that forms the framework for the system (depending on the model eg. with display, PC communication interfaces, storage meory options etc.). It offers several slots that can be equipped with functional I/O units. These I/O modules define the I/O channels of the system and with this determin the measurement functions. Several units can be combined to offer exactly the channels required for individual applications.
Note, that not all modules can be used in all man units. Moreover, in some devices certain combinations of units may be not supported (see datasheets for detailed information).
HIOKI U8976, 8966, U8975, U8978, 8968, 8972, U8974 General Voltage Modules
Measured signal | Model | Description | Compatible with | Number of channels | Fastest sampling | Bandwidth | A/D resolution | DC accuracy (f.s.) | Max. input voltage | Sensitivity1) | Max. sensitivity range (f.s.) | Isolation | Supplement |
Voltage (high speed) | U8976 | High-speed analog unit | MR6000 | 2 | 200 MS/s | DC...30 MHz | 12 bit | ±0.5% | 400 VDC, 1000 VDC2) | 0.0625 mV | 100 mV | Yes | - |
Voltage | 8966 | Analog unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 20 MS/s | DC...5 MHz | 12 bit | ±0.5% | 400 VDC | 0.05 mV | 100 mV | Yes | - |
Voltage (4-channel) | U8975 | 4-channel analog unit | MR6000, MR8847 | 4 | 5 MS/s | DC...2 MHz | 16 bit | ±0.1% | 200 VDC | 0.125 mV | 4 V | Yes | - |
Voltage (4-channel, high resolution) | U8978 | 4-channel analog unit | MR6000, MR8847 | 4 | 5 MS/s | DC...2 MHz | 16 bit | ±0.3% | 40 VDC | 3.125 µV | 100 mV | Yes | - |
Voltage (high resolution) | 8968 | High resolution unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 1 MS/s | DC...100 kHz | 16 bit | ±0.3% | 400 VDC | 3.125 µV | 100 mV | Yes | with AAF |
Voltage (DC, RMS) | 8972 | DC/RMS unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 1 MS/s | DC...400 kHz | 12 bit | ±0.5% | 400 VDC | 0.05 mV | 100 mV | Yes | with RMS |
Voltage (high voltage) | U8974 | High voltage unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 1 MS/s | DC...100 kHz | 16 bit | ±0.25% | 1000 VDC; 700 VAC | 0.125 mV | 4 V | Yes | - |
HIOKI MR8990, U8991 Voltmeter Modules
Measured signal | Model | Description | Compatible with | Number of channels | Fastest sampling | Bandwidth | A/D resolution | DC accuracy (f.s.) | Max. input voltage | Sensitivity1) | Max. sensitivity range (f.s.) | Isolation | Supplement |
Voltage (high resolution) | MR8990 | Digital voltmeter unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 2 ms | - | 24 bit | ±0.01% rdg. ±0.0025% f.s. | 500 VDC | 0.1 µV | 100 mV | Yes | - |
Voltage | U8991 | 6½-digit, 1,200,000 counts digital voltmeter unit | MR8740T | 4 | 50 times/s | DC voltage measurement; 3 ranges 1, 10, 100 V; resolution 1/1,000,000 of measurement range (using 24-bit ΔΣ modulation A/D); basic measurement accuracy ±0.02% rdg. ±0.0025% f.s.; max. input voltage 100 VDC | Yes | - |
HIOKI U8977, 8971 Current Modules
Measured signal | Model | Description | Compatible with | Number of channels | Fastest sampling | Bandwidth | A/D resolution | DC accuracy (f.s.) | Max. input voltage | Sensitivity1) | Max. sensitivity range (f.s.) | Isolation | Supplement |
Current | U8977 | 3-channel current unit | MR6000, MR8847 | 3 | 5 MS/s | DC...2 MHz | 16 bit | ±0.3% | Current sensor only | Depends on current sensor | - | Max. 3 units | |
Current | 8971 | Current unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 1 MS/s | DC...100 kHz | 12 bit | ±0.65% | Current sensor only | Depends on current sensor | - | with RMS; max. 4 units |
HIOKI 8967, U8969, 8970, U8979 Sensor and Frequency Modules
Measured signal | Model | Description | Compatible with | Number of channels | Fastest sampling | Bandwidth | A/D resolution | DC accuracy (f.s.) | Max. input voltage | Sensitivity1) | Max. sensitivity range (f.s.) | Isolation | Supplement |
Temperature | 8967 | Temperature unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 1.2 ms | DC | 16 bit | see detailed reference | Thermocouples only | 0.01°C | 200°C (392°F) | Yes | - |
Strain | U8969 | Strain unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 200 kS/s | DC...20 kHz | 16 bit | ±0.5% ±4 µε | Strain only | 0.016 µε | 400 µε | Yes | Discontinued product 8969 can also be used |
Frequency | 8970 | Frequency unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | 200 kS/s | DC...100 kHz3) | 16 bit | - | 400 VDC | 0.002 Hz | Depends on mode | Yes | - |
Acceleration | U8979 | Charge unit | MR6000, MR8847 | 2 | 200 kS/s | DC...50 kHz (DC); 1 Hz...50 kHz (AC) | 16 bit | ±0.5% (voltage); ±2.0% (acceleration) | 40 VDC | Depends on acceleration sensor | Yes | Supports TEDS |
HIOKI 8973 Logic Module
Measured signal | Model | Description | Compatible with | Number of channels | Fastest sampling | Bandwidth | A/D resolution | DC accuracy (f.s.) | Max. input voltage | Sensitivity1) | Max. sensitivity range (f.s.) | Isolation | Supplement |
Logic | 8973 | Logic unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 4 probes (16 channels) | - | - | - | - | - | - | - | - | Requires 9320-01, 9327 or MR9321-01 |
1) Minimum resolution shows the highest sensitivity resolution.
2) When using the 9665.
3) Minimum pulse width 2 µs.
HIOKI U8793, MR8790, MR8791, U8794 Generator and Output Modules
Output signal | Model | Description | Compatible with | Number of channels | Output function | Output voltage range | |||
Waveform generation | U8793 | Arbitrary waveform generator unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 2 | Function generator: Sine, square, pulse, triangle, ramp, DC; arbitrary waveform generation: Measurement waveform with memory recorder, waveform editted with the SP8000 | -10...15 V | |||
Waveform generation | MR8790 | Waveform generator unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 4 | DC, sine wave (output frequency range: 0 Hz...20 kHz) | -10...10 V | |||
Pulse generation | MR8791 | Pulse generator unit | MR6000, MR8740(T), MR8741, MR8827, MR8847 | 8 | Pulse output: Frequency 0 Hz...20 kHz; logic output: Voltage level 0...5 V, open-collector | Output terminal connector: D-sub, half-pitch, 50-pin | |||
DC voltage, DC current, resistance simulation output | U8794 | VIR generator unit | MR8740T | 8 | DC voltage: -0.1000...+5.3000 V (setting resolution: 0.1 mV); DC current: 5 mA range: -5.0000...+5.0000 mA (setting resolution: 0.1 µA); 1 mA range: -1.00000...+1.00000 mA (setting resolution: 0.01 µA); 250 µA range: -250.00...+250.00 µA (setting resolution: 0.01 µA); 50 µA range: -50.000...+50.000 µA (setting resolution: 0.001 µA); Resistance: 10 Ω...1 MΩ (setting resolution: 6 digits) Output accuracy: DC voltage: 5 V range, ±0.035% of setting ±800 µV; DC current: 5-mA range: ±0.050% of setting ±4.0 µA; 1-mA range: ±0.050% of setting ±800 nA; 250-µA range: ±0.050% of setting ±200 nA; 50-µA range: ±0.050% of setting ±40 nA |